Related Experiment Video
Updated: Jul 30, 2025

Treatment of Ankle Osteoarthritis with Total Ankle Replacement Through a Lateral Transfibular Approach
Published on: January 24, 2018
[Biomechanics of the first metatarsophalangeal joint after replacement]
A V Kolsanov1, A N Nikolaenko1, V V Ivanov1
1Samara State Medical University, Samara, Russia.
Objective:
Analysis of biomechanics of the first metatarsophalangeal joint after arthroplasty, interaction between bones and two implants of the first metatarsophalangeal joint using skeletal model of the foot.
Material And Methods:
We developed anatomically adapted all-ceramic non-coupled endoprosthesis of proximal interphalangeal joint between 2016 and 2021. To create a model of the foot, we used diagnostic computed tomography whose images were applied in 3D sculpting system and computer-aided design system for final geometric modeling of the joint.
Results:
In dorsal flexion of the first metatarsophalangeal joint under 45° with the presence of implant, cortical bone tissue can withstand a load of up to 40 kg. Cortical bone tissue with implant can withstand a load of up to 305 kg without dorsal flexion. Strength of implant elements made of zirconium ceramics significantly exceeds strength of bone tissue within implant-bone tissue connection.
Conclusion:
Postoperative axial load on the first metatarsophalangeal joint up to 35 kg with maximum dorsal flexion up to 45° is the most appropriate. Higher load and hyperextension over 45° may be followed by postoperative complications such as implant instability, dislocation and periprosthetic fracture.