Technical and clinical assessment of latest technology SiPM integrated digital PETCT scanner
Objective:
The aim of this study was to conduct a technical and clinical evaluation of a Silicon Photomultiplier (SiPM) integrated digital Positron Emission Tomography - Computed Tomography (PETCT) Scanner using National Electrical Manufacturers Association (NEMA) NU 2- 2018 standards.
Methods:
System sensitivity was measured by using a NEMA sensitivity phantom. Scatter fraction, count-rate performance, accuracy of count loss, and timing resolution were all computed. Clinical images were acquired and image quality was assessed and compared with published studies.
Results:
At 1 cm, tangential, radial, and axial spatial resolutions were 3.02 mm, 3.02 mm, and 2.73 mm at full width half maximum (FWHM), respectively. Sensitivity at centre and 10 cm was 10.359 cps/kBq and 9.741 cps/kBq, respectively. The timing resolution was measured at 372 ps.
Conclusion:
The digital PETCT exhibits a high-spatial resolution and a superior timing resolution, which advances the diagnostic ability to detect small lesions and boosts the diagnostic confidence.
Implications For Practice:
Increases clinical relevance by improving the ability to detect and differentiate tiny or low-contrast lesions without compromising radiopharmaceutical dose or overall scan time.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography