Related Experiment Video
Updated: Jul 30, 2025

Purification of Platelets from Mouse Blood
Published on: May 7, 2019
Platelet count in impedance-based hematology analyzer: Beware of trap!
Vani Krishnamurthy1, Archana Shivamurthy1, Pradeep V Kumar2
1Department of Pathology, JSS Medical College, Mysore, Karnataka, India.
Abstract:
Among the blood counts, platelet count is most often reported with inconsistency. Many of the analyzers work on electrical impedance principle for red blood cell (RBC) and platelet counting. However, with this technology, factors such as fragmented RBCs, microcytes, cytoplasmic fragments of leukemic cells, lipid particles, fungal yeast forms, and bacteria are known to interfere with platelet count and give spuriously elevated platelet counts. A 72-year-old male was admitted for the treatment of dengue infection who had serial platelet count monitoring. He had an initial platelet count of 48,000/cumm which suddenly improved to 2.6 lakhs within 6 h without any platelet transfusion. Peripheral smear however did not correlate with the machine-derived count. Repeat test after 6 h yielded a result of 56,000/cumm which correlated well with the peripheral smear. This falsely elevated count was due to the presence of lipid particles as the sample was drawn in the postprandial state.
More Related Videos
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
10:25Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro
Published on: March 19, 2016