Related Experiment Video
Updated: Jul 30, 2025

Author Spotlight: Improving Radiation Therapy Access with Radiation Planning Assistant
Published on: October 6, 2023
A virtual audit system for intensity-modulated radiation therapy credentialing in Japan Clinical Oncology Group
Mitsuhiro Nakamura1, Dejun Zhou1, Toshiyuki Minemura2
1Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
Purpose:
The Medical Physics Working Group of the Radiation Therapy Study Group at the Japan Clinical Oncology Group is currently developing a virtual audit system for intensity-modulated radiation therapy dosimetry credentialing. The target dosimeters include films and array detectors, such as ArcCHECK (Sun Nuclear Corporation, Melbourne, Florida, USA) and Delta4 (ScandiDos, Uppsala, Sweden). This pilot study investigated the feasibility of our virtual audit system using previously acquired data.
Methods:
We analyzed 46 films (32 and 14 in the axial and coronal planes, respectively) from 29 institutions. Global gamma analysis between measured and planned dose distributions used the following settings: 3%/3 mm criteria (the dose denominator was 2 Gy), 30% threshold dose, no scaling of the datasets, and 90% tolerance level. In addition, 21 datasets from nine institutions were obtained for array evaluation. Five institutions used ArcCHECK, while the others used Delta4. Global gamma analysis was performed with 3%/2 mm criteria (the dose denominator was the maximum calculated dose), 10% threshold dose, and 95% tolerance level. The film calibration and gamma analysis were conducted with in-house software developed using Python (version 3.9.2).
Results:
The means ± standard deviations of the gamma passing rates were 99.4 ± 1.5% (range, 92.8%-100%) and 99.2 ± 1.0% (range, 97.0%-100%) in the film and array evaluations, respectively.
Conclusion:
This pilot study demonstrated the feasibility of virtual audits. The proposed virtual audit system will contribute to more efficient, cheaper, and more rapid trial credentialing than on-site and postal audits; however, the limitations should be considered when operating our virtual audit system.
Related Concept Videos
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Integrated Healthcare System
Absorption of Radiation
Nursing Clinical Information System
A Nursing Clinical Information System (NCIS) is a specialized type of healthcare information system tailored to meet the unique needs of nursing practice. It incorporates the principles of nursing informatics to streamline information management and improve the quality of care delivery.
Critical attributes of NCIS include:
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...