Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
3D-Ring CZT System With New Low- and Medium-Energy Range: Ultrafast Dual-Isotope Lung SPECT/CT Improvement
Aurélien Callaud1, Gilles Metrard, Matthieu Bailly
1From the Nuclear Medicine Department, CHU ORLEANS, Orleans, France.
Abstract:
New 3D-ring CZT systems with low- and medium-energy-range detectors allow for simultaneous dual-isotope lung scintigraphy. We compared 10-, 7-, 5-, and 3-minute acquisitions for 99m Tc and 81m Kr performed simultaneously on StarGuide CZT-SPECT/CT and reframed in 50 patients. Ventilation/perfusion mismatches were calculated (mean, 15.6% ± 28%), and Spearman correlation coefficients of mismatches were 0.994, 0.994, and 0.984 between 10- and 7-, 5-, and 3-minute acquisitions, respectively. No visual difference in image quality or final diagnosis was found. 3D-ring CZT-SPECT with low and medium energy range detectors allows ultrafast dual-isotope lung scintigraphy up to 3 minutes.