Artificial Lung Device Priming for In Situ Fiber Bundle Surface Grafting
Published on: March 28, 2025
Comparison of complement consumption and platelet accumulation between membrane oxygenators coated with a polymer or
Masashi Tagaya1, Takuo Murataka1, Shinya Okano1
1Department of Medical Engineering, National Hospital Organization Kure Medical Center and Chugoku Cancer Center, Hiroshima, Japan.
Introduction:
The membrane oxygenator in extracorporeal circulation circuits is coated with acrylate-copolymer (ACP) or immobilized heparin (IHP) to enhance hemocompatibility. To evaluate the relative features of both coatings, we compared blood components circulated in the circuits with ACP-and IHP-coated membranes in vitro using whole human blood.
Methods:
Whole human blood was heparinized and circulated in two experimental circuits with an ACP-coated reservoir, tubes, and an ACP- or IHP-coated membrane. Platelet (PLT) counts and the amount of total protein (TP), complement component 3 (C3), and complement component 4 (C4) were measured at 0, 8, 16, 24, and 32 h in each experiment (n = 5).
Results:
The PLT count at 0-h circulation was lower in the IHP-coated than in the ACP-coated circuits (p = 0.034); however, no significant difference was observed at other time points. Reduction in TP at 8-h and 16-h circulation and in C3 at 32-h circulation was lesser in the ACP-coated than in the IHP-coated circuits (p = 0.004, 0.034, and 0.027, respectively); reduction in TP and C3 at other time points and C4 at each time point was not significantly different. There were significant interactions between coating type and circulation duration in the PLT, TP, and C3 transitions (p = 0.008, 0.020, and 0.043, respectively).
Conclusions:
Our findings suggest that ACP-coated membranes can prevent the initial drop in PLT count and C3 consumption over 32 h, whereas IHP-coated membranes could not prevent this drop in extracorporeal circulation. Therefore, ACP-coated membranes are suitable for short- and long-term extracorporeal life support.
More Related Videos
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
Related Concept Videos
Formation of the Platelet Plug
As the injured blood vessel contracts, endothelial cells undergo contraction, revealing collagen fibers in the basement membrane and underlying connective tissue. Furthermore, the plasma membrane of endothelial cells becomes adhesive, preparing the site for platelet adhesion. Platelets...
Anticoagulant Drugs: Low-Molecular-Weight Heparins