Related Experiment Video
Updated: Jul 29, 2025

A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023
A radiologic determination of the different screw cutting patterns in cut and uncut orthopedic cortical screws using
Joshua W Major1, Andrew J Ernst2, Jonathan K Kallevang2
1Department of Radiology, Naval Medical Center Portsmouth, Portsmouth, VA, 23708-2197, USA. jmajor08@gmail.com.
Objective:
We hypothesize that cut screws will deform in a manner that increases the core and outer diameters of the screw hole compared to uncut controls, and effects will be more pronounced in titanium screws.
Materials And Methods:
We used biomechanical polyurethane foam blocks to simulate cortical bone. We organized four groups of stainless steel and titanium cut and uncut screws. Blocks were fitted with a jig to ensure perpendicular screw insertion. We imaged the blocks using digital mammography and measured them using PACS software. Power analysis determined a power of 0.95 and an alpha error of 0.05.
Results:
Highly statistically significant differences in core diameter were found after cutting stainless steel and titanium screws. Cutting stainless steel screws increased core diameter by 0.30 mm (95% CI, 0.16 to 0.45; p < .001). Titanium screws' core diameter increased by 0.45 mm (95% CI, 0.30 to 0.61; p < .001). No significant differences were found in the outer diameters of stainless steel and titanium screws after cutting.
Conclusion:
Titanium and stainless steel screw tracts demonstrated screw core diameter and screw thread pattern deformation after cutting. Titanium screws demonstrated more significant effects.