Related Experiment Video
Updated: Jul 28, 2025

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Dosimetric Study to Estimate Deviations in Delivered Radiation Dose due to Occluded Air Spaces in Vaginal Vault
Challapalli Srinivas1, Sourjya Banerjee1, Dilson Lobo1
1Department of Radiation Oncology, Kasturba Medical College, Mangalore , Manipal Academy of Higher Education, Manipal, India.
Background:
To develop a dosimetric tool to estimate the dose delivered in the presence of air pockets with EBT3 film while simulating the conditions of vaginal vault brachytherapy (VVBT) with 3.0 diameter cylindrical applicator at a prescription dose distance of 5mm from the surface of it.
Materials And Method:
Six acrylic plates (10 cm x 10 cm, 0.5 cm thick) with four different types of slots were designed and produced locally. They can hold a cylindrical vaginal brachytherapy applicator in the centre, air equivalent material from the applicator's surface [(sizes 4.5 mm (A), 3.0 mm (B), and 2.0 mm (C)], EBT3 film at the prescribed dose distance, and holder rods. Plates were layered together with acrylic rods and assembled in a holding box in a water phantom. Three treatment plans done in TPS with prescription doses of 2 Gy, 3 Gy, and 4 Gy at 5.0 mm with a treatment length of 6 cm, and were executed in Co-60-based HDR brachytherapy unit (M/s SagiNova, Germany) with & without the placement of air equivalent material, and the dose received at slot locations A, B, & C were noted.
Results:
The mean percentage deviation of measured dose without and with presence of air pocket at A, B and C was 13.9%, 11.0% and 6.4% respectively for all dose prescriptions. As the air pocket size expanded radially from 2.0 mm to 4.5 mm, the increase in dosage ranged from 6.4% to 13.9% which was due to the fact that the film was held at dosage prescription distance and the lack of attenuation of photons radially through air pocket.
Conclusions:
The present study can be carried out with a 3D printed phantom that simulates VVBT application having air pockets of different dimensions at different locations and also can be analyzed with Monte Carlo simulations.
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Absorption of Radiation
Biological Effects of Radiation