In Vivo Evaluation of the Mechanical and Viscoelastic Properties of the Rat Tongue
Published on: July 6, 2017
Concurrent Validity of the IOPI and Tongueometer Orofacial Strength Measurement Devices
James A Curtis1,2, Valentina Mocchetti1, Anaïs Rameau1,3
1Department of Otolaryngology-Head & Neck Surgery, Weill Cornell Medicine, Sean Parker Institute for the Voice, New York, NY, USA.
Objective(S):
This study examined the concurrent validity of two orofacial strength manometers: (1) the Iowa Oral Performance Instrument (IOPI) - the current, gold standard orofacial manometer; and (2) the Tongueometer - a newly-available, lower cost, orofacial manometer.
Methods:
This study compared IOPI and Tongueometer pressure readings across three experimental conditions. Experiment 1 compared full setup (manometer + tongue bulb) pressure readings between the IOPI and Tongueometer. Experiment 2 compared IOPI tongue bulb and Tongueometer tongue bulb pressure readings, while controlling for manometer. Experiment 3 compared IOPI manometer and Tongueometer manometer pressure readings, while controlling for tongue bulb. Pressures were applied manually within a laboratory setting. Lin's concordance correlation (ρc ) was used to calculate level of agreement, with ρc interpreted as 'poor' if <0.90, 'moderate' if 0.90 to <0.95, 'substantial' if 0.95 to <0.99, and 'excellent' if ≥0.99.
Results:
539 trials were analyzed. There was a median absolute difference of 2.4 kPa in pressure readings between the IOPI and Tongueometer full setups (manometer + tongue bulb). Correlations revealed substantial agreement between IOPI and Tongueometer full setups (experiment 1: n = 292; ρc = 0.986), tongue bulbs (experiment 2: n = 146; ρc = 0.987-0.992), and manometers (experiment 3: n = 101; ρc = 0.970).
Conclusions:
Differences in pressures were consistently observed between the Tongueometer and IOPI. Despite these differences, substantial agreement was present. These data suggest the Tongueometer may be a valid, lower cost alternative to the IOPI for objectively assessing orofacial strength in clinical practice.
Level Of Evidence:
Level 2 Laryngoscope, 133:3123-3131, 2023.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
06:00Assessment of Spatial Lingual Tactile Sensitivity using a Gratings Orientation Test
Published on: September 17, 2021
Related Concept Videos
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...