Related Experiment Video
Updated: Jul 28, 2025

Oral Biofilm Analysis of Palatal Expanders by Fluorescence In-Situ Hybridization and Confocal Laser Scanning Microscopy
Published on: October 20, 2011
Identification of Pseudomonas aeruginosa exopolysaccharide Psl in biofilms using 3D OrbiSIMS
Heba Khateb1,2, Andrew L Hook1, Stefanie Kern3
1Advanced Materials and Healthcare Technologies Division, School of Pharmacy, University of Nottingham, Nottingham, United Kingdom.
Abstract:
Secondary ion mass spectrometry (SIMS) offers advantages over both liquid extraction mass spectrometry and matrix assisted laser desorption mass spectrometry in that it provides the direct in situ analysis of molecules and has the potential to preserve the 3D location of an analyte in a sample. Polysaccharides are recognized as challenging analytes in the mass spectrometry of liquids and are also difficult to identify and assign using SIMS. Psl is an exopolysaccharide produced by Pseudomonas aeruginosa, which plays a key role in biofilm formation and maturation. In this Letter, we describe the use of the OrbiTrap analyzer with SIMS (3D OrbiSIMS) for the label-free mass spectrometry of Psl, taking advantage of its high mass resolving power for accurate secondary ion assignment. We study a P. aeruginosa biofilm and compare it with purified Psl to enable the assignment of secondary ions specific to the Psl structure. This resulted in the identification of 17 peaks that could confidently be ascribed to Psl fragments within the biofilm matrix. The complementary approach of the following neutral loss sequences is also shown to identify multiple oligosaccharide fragments without the requirement of a biological reference sample.
More Related Videos
07:25Visualization of Pseudomonas aeruginosa within the Sputum of Cystic Fibrosis Patients
Published on: July 16, 2020
11:09An Analytical Tool-box for Comprehensive Biochemical, Structural and Transcriptome Evaluation of Oral Biofilms Mediated by Mutans Streptococci
Published on: January 25, 2011