Limitations of transcutaneous carbon dioxide monitoring in apneic oxygenation
Thilo Schweizer1, Volker Hartwich2, Thomas Riva1,3
1Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
Background:
High-flow nasal oxygenation is increasingly used during sedation procedures and general anesthesia in apneic patients. Transcutaneous CO2 (ptcCO2)-monitoring is used to monitor hypercapnia. This study investigated ptcCO2-monitoring during apneic oxygenation.
Methods:
We included 100 patients scheduled for elective surgery under general anesthesia in this secondary analysis of a randomized controlled trial. Before surgery, we collected ptcCO2 measured by TCM4 and TCM5 monitors and arterial blood gas (ABG) measurements every two minutes during 15 minutes of apnea. Bland-Altman plots analyzed agreement between measurement slopes; linear mixed models estimated the different measuring method effect, and outlined differences in slope and offset between transcutaneous and arterial CO2 partial pressures.
Results:
Bland-Altman plots showed a bias in slope (95% confidence intervals) between ABG and TCM4-measurements of 0.05mmHg/min (-0.05 to 0.15), and limits of agreement were -0.88mmHg/min (-1.06 to -0.70) and 0.98mmHg/min (0.81 to 1.16). Bias between ABG and TCM5 was -0.14mmHg/min (-0.23 to -0.04), and limits of agreement were -0.98mmHg/min (-1.14 to -0.83) and 0.71mmHg/min (0.55 to 0.87). A linear mixed model (predicting the CO2-values) showed an offset between arterial and transcutaneous measurements of TCM4 (-15.2mmHg, 95%CI: -16.3 to -14.2) and TCM5 (-19.1mmHg, -20.1 to -18.0). Differences between the two transcutaneous measurements were statistically significant.
Conclusions:
Substantial differences were found between the two transcutaneous measurement systems, and between them and ABG. Transcutaneous CO2 monitoring cannot replace arterial CO2-monitoring during apneic oxygenation. In clinical settings with rapidly changing CO2-values, arterial blood gas measurements are needed to reliably assess the CO2-partial pressure in blood.
Trial Registration:
ClinicalTrials.gov (NCT03478774).
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-III
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: