Assessment of cardiac load-responsiveness in veno-arterial extracorporeal life support: A case series
Y Cornelisse1, P W Weerwind1, M E Bol2,3
1Department of Extra-Corporeal Circulation and Cardiothoracic Surgery, Maastricht University Medical Centre, Maastricht, Netherlands.
Introduction:
Well-timed explant of veno-arterial extracorporeal life support (V-A ECLS) depends on adequate assessment of cardiac recovery. Often, evaluation of cardiac recovery consists of reducing support flow while visualizing cardiac response using transoesophageal echocardiography (TEE). This method, however, is time consuming and based on subjective findings. The dynamic filling index (DFI) may aid in the quantitative assessment of cardiac load-responsiveness. The dynamic filling index is based on the relationship of support flow and pump speed, which varies with varying hemodynamic conditions. This case series intends to investigate whether the DFI may support TEE in facilitating the assessment of cardiac load-responsiveness.
Methods:
Measurements for DFI-determination were performed in seven patients while simultaneously assessing ventricular function by measuring the aortic velocity time integral (VTI) using TEE. Measurements consisted of multiple consecutive transient speed manipulations (∼100 r/min) during weaning trials, both at full support and during cardiac reloading at reduced support.
Results:
The VTI increased between full and reduced support in six weaning trials. In five of these trials DFI decreased or remained equal, and in one case DFI increased. Of the three trials in which VTI decreased between full and reduced support, DFI increased in two cases and decreased in one case. Changes in DFI, however, are mostly smaller than the detection threshold of 0.4 mL/rotation.
Conclusion:
Even though current level of accuracy of the parameter requires further investigation to increase reliability and possibly predictability, DFI seems likely to be a potential parameter in supporting TEE for the assessment of cardiac load-responsiveness.
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...