Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
Published on: February 20, 2020
Intra- and inter-rater reliability, agreement, and minimal detectable change of the handheld dynamometer in
Gilvan Ferreira Vaz1,2, Felipe Florêncio Freire2, Henrique Mansur Gonçalves3
1Faculty of Ceilândia, Rehabilitation Sciences Program, University of Brasilia, Brasília, DF, Brazil.
Introduction:
The handheld dynamometer has been validated to measure muscle strength in different muscle groups. However, to date, it has not been tested in individuals who experience pain induced by hip osteoarthritis. The current study aimed to evaluate the intra- and inter-rater reliability, agreement, and minimal detectable change of the Lafayette handheld dynamometer, model 1165, to assess the peak force (Pk) and average peak force (Af) of hip muscles in individuals with symptomatic hip osteoarthritis.
Methods:
Twenty participants with hip osteoarthritis (mean ± SD age: 58.7±15.3 years; body mass index: 28.8±4.2 kg/m2) and pain intensity on the Visual Analogue Scale ≥ 4 (8.05±1.2) were recruited to participate in this study. Pk and Af of hip flexors (seated position), abductors and adductors (supine position), and extensors (prone position) were collected in a single day by two independent raters, each one obtaining test and retest in randomly ordered separate sessions.
Results:
The intra-rater intraclass correlation coefficient (ICC) was classified as good (>0.75) or excellent (≥0.90) for all muscle groups and all inter-rater ICCs were classified as excellent. Rater A had a lower standard error of measurement compared to rater B, ranging from 0.15 to 0.58 kilogram-force (Kgf) compared with 0.34 to 1.25 kg, respectively. However, the inter-rater comparison showed a minimal detectable change (MDC) of < 10% for all Pk and Af measures for hip adductors and extensors. Finally, the inter-rater Bland-Altman analysis demonstrated good agreement for abductors, adductors, and extensors.
Conclusion:
Despite pain and dysfunction related to hip osteoarthritis, the mean of two measures using a handheld dynamometer was shown to be a reliable tool to assess hip muscle strength, with good to excellent intra- and inter-rater ICCs, satisfactory agreement, and small values for MDC.
More Related Videos
07:43In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
07:22Glycemic Impact on Knee Osteoarthritis Symptoms on Physical, Radiographic, and Inflammatory Markers among Individuals Aged 50 and Over with Diabetes
Published on: March 7, 2025
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...