Posterior vitreous fluorophotometry in normal subjects
Archives of Ophthalmology (Chicago, Ill. : 1960)
|July 1, 1986
Summary
Vitreous fluorophotometry in normal subjects established reference ranges for fluorescein concentration and permeability. These findings help identify abnormally high fluorescein leakage in the posterior vitreous.
Area of Science:
- Ophthalmology
- Biomedical Engineering
Background:
- Vitreous fluorophotometry is a key technique for assessing ocular barrier function.
- Establishing normal reference ranges is crucial for diagnosing ocular pathologies involving leakage.
Purpose of the Study:
- To determine normal vitreous fluorescein concentration and permeability index values in healthy subjects.
- To establish criteria for identifying abnormally high fluorescein leakage in the posterior vitreous.
Main Methods:
- Vitreous fluorophotometry was performed using the Fluorotron Master after intravenous fluorescein injection (14 mg/kg) in 22 normal subjects.
- Fluorescein concentration was measured at 3 mm from the retina and averaged over the posterior 6 mm.
- Permeability index was calculated 60 minutes post-injection.
Main Results:
- Mean post-injection vitreous fluorescein concentrations ranged from 1.7-4.3 ng/mL at 30 minutes and 7.1-10.8 ng/mL at 60 minutes.
- The permeability index ranged from 0.92-1.19 X 10(-7) cm/s.
- Reproducibility of measurements varied from 17% to 49% depending on the analysis method.
Conclusions:
- Normal reference ranges for posterior vitreous fluorescein concentration and permeability were established.
- Abnormally high fluorescein leakage may be indicated by a 60-minute vitreous concentration >14.3 ng/mL or a permeability index >2 X 10(-7) cm/s.