Background:
With the recent developments in automated tools, smaller and cheaper machines for lung ultrasound (LUS) are leading us toward the potential to conduct POCUS tele-guidance for the early detection of pulmonary congestion. This study aims to evaluate the feasibility and accuracy of a self-lung ultrasound study conducted by hemodialysis (HD) patients to detect pulmonary congestion, with and without artificial intelligence (AI)-based automatic tools.
Methods:
This prospective pilot study was conducted between November 2020 and September 2021. Nineteen chronic HD patients were enrolled in the Soroka University Medical Center (SUMC) Dialysis Clinic. First, we examined the patient's ability to obtain a self-lung US. Then, we used interrater reliability (IRR) to compare the self-detection results reported by the patients to the observation of POCUS experts and an ultrasound (US) machine with an AI-based automatic B-line counting tool. All the videos were reviewed by a specialist blinded to the performer. We examined their agreement degree using the weighted Cohen's kappa (Kw) index.
Results:
A total of 19 patients were included in our analysis. We found moderate to substantial agreement between the POCUS expert review and the automatic counting both when the patient performed the LUS (Kw = 0.49 [95% CI: 0.05-0.93]) and when the researcher performed it (Kw = 0.67 [95% CI: 0.67-0.67]). Patients were able to place the probe in the correct position and present a lung image well even weeks from the teaching session, but did not show good abilities in correctly saving or counting B-lines compared to an expert or an automatic counting tool.
Conclusions:
Our results suggest that LUS self-monitoring for pulmonary congestion can be a reliable option if the patient's count is combined with an AI application for the B-line count. This study provides insight into the possibility of utilizing home US devices to detect pulmonary congestion, enabling patients to have a more active role in their health care.
More Related Videos
06:15Protocol and Guidelines for Point-of-Care Lung Ultrasound in Diagnosing Neonatal Pulmonary Diseases Based on International Expert Consensus
Published on: March 6, 2019
03:47Author Spotlight: Workflow for Integrating POCUS Data into EHR for Managing Heart Failure Patients
Published on: July 12, 2024
Related Concept Videos
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Imaging Studies II: Ultrasonography