Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger
Jorge Herranz Olazabal1,2, Ilde Lorato1, Jesse Kling1
1IMEC NL, 5656 AE Eindhoven, The Netherlands.
Abstract:
Speckle Plethysmography (SPG) and Photoplethysmography (PPG) are different biophotonics technologies that allow for measurement of haemodynamics. As the difference between SPG and PPG under low perfusion conditions is not fully understood, a Cold Pressor Test (CPT-60 s full hand immersion in ice water), was used to modulate blood pressure and peripheral circulation. A custom-built setup simultaneously derived SPG and PPG from the same video streams at two wavelengths (639 nm and 850 nm). SPG and PPG were measured at the right index finger location before and during the CPT using finger Arterial Pressure (fiAP) as a reference. The effect of the CPT on the Alternating Component amplitude (AC) and Signal-to-Noise Ratio (SNR) of dual-wavelength SPG and PPG signals was analysed across participants. Furthermore, waveform differences between SPG, PPG, and fiAP based on frequency harmonic ratios were analysed for each subject (n = 10). Both PPG and SPG at 850 nm show a significant reduction during the CPT in both AC and SNR. However, SPG showed significantly higher and more stable SNR than PPG in both study phases. Harmonic ratios were found substantially higher in SPG than PPG. Therefore, in low perfusion conditions, SPG seems to offer a more robust pulse wave monitoring with higher harmonic ratios than PPG.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: