Tension controlled hollow-fiber winding machine for blood oxygenator prototypes
Easa AliAbbasi1, Anil Akseki2, Azmat Ullah3
1Department of Computational Sciences and Engineering, Koc University, Istanbul, Turkey.
Abstract:
Blood oxygenators involve a complex network of hollow fibers for efficient gas exchange with blood. The optimal microstructural arrangement of these fibers is an ongoing research interest. While the fiber systems of commercial oxygenators are manufactured to address mass production, the research oxygenator prototypes demand more flexibility so that different design parameters can be tested. Here a hollow-fiber assembly system is designed and built for winding research grade extracorporeal blood oxygenator mandrels at different layout dimensions so that these different configurations can be evaluated for mass transfer capacity and blood damage. The hardware design and manufacturing details of this system presented together with its impact on the prototype oxygenator device assembly process. This in-house built system can wind thin fibers, having outer diameters ranging from 100 μm to 1 mm, at any specified winding angle continuously. A control system for fiber stress is also incorporated to eliminate fiber damage. Our system consists of three main units: (1) unwinding, (2) accumulator, and (3) winding systems, integrated together via the control software. The unwinding unit has a PID controller to maintain the position of the accumulator motor on the reference point by tuning the velocity of feeding fibers to the accumulator unit. Another PID controller preserves the desired tension value of the fibers by adjusting the position of the accumulator motor. Desired tension value is defined by the user and typically obtained through uniaxial testing of fibers. The control unit employs a "cascaded" PID controller since the PID controller in the accumulator unit maintains the tension and the PID controller in the unwinding unit controls the position of the accumulator motor. Finally, the winding unit utilizes two motors to wind the fibers over the outer diameter of a mandrel at the desired winding angle. The first motor drives the translational movement, and the second one provides mandrel rotation. The desired angles are achieved by tuning the synchronous movement of the winding motors. While the system is designed to produce assembled blood oxygenator mandrel prototypes, this concept is also applicable for producing cylindrical fiber-reinforced composite materials with specified fiber angles and stents winded on jigs.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...