Related Experiment Video
Updated: Jul 26, 2025

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Detection of atrial fibrillation in real world setting in patients with cryptogenic stroke and an implantable loop
Seulgi Erica Kim1, Muzamil Khawaja1, Jitae Alex Kim1
1Department of Internal Medicine, Baylor College of Medicine, Houston, Texas, USA.
Introduction:
Implantable loop recorders (ILR) are used to screen for atrial fibrillation (AF) in patients with cryptogenic stroke (CS). However, there is limited real-world data regarding the long-term rate of AF detection using ILR and management consequences in patients with CS. The objective is to assess the rate of AF detection in patients with CS in a real-world study over 36 months of follow-up and its consequences on stroke prevention.
Methods:
This retrospective study included patients with an ILR placed for CS at Baylor College of Medicine and Baylor St. Luke's Medical Center between January 2014 and July 2021. The primary outcome was AF detection in patients with ILR. The secondary outcome was the rate of subsequent strokes after ILR placement in patients with or without diagnosed AF. The AF detection rate in our cohort was compared to the rate in CRYSTAL-AF Trial at 36-month follow-up. The impact of AF detection on clinical management was examined.
Results:
We identified 225 patients. 51.1% were women and 38.2% African American. Among 85 patients with ILR labeled AF, 43 patients had true AF, and 42 had incorrectly labeled AF (48.3% false positive). The estimated AF detection rate at 36 months follow-up was 28.6% (95% CI, 26.6%-30.6%). 58.1% of patients with AF were initiated on oral anticoagulation, 80.0% of whom were started on a direct oral anticoagulant. 13.8% of patients had recurrent strokes after ILR implantation; 4 of whom were diagnosed with AF.
Conclusion:
Compared to CRYSTAL-AF, the AF detection rate in our cohort is similar, but this cohort includes a higher proportion of female and African American patients. Most patients with recurrent strokes after ILR implant did not have AF during 36 months of monitoring.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Acute Coronary Syndrome III: Diagnostic Studies