Fluorescence-quenching of a Liposomal-encapsulated Near-infrared Fluorophore as a Tool for In Vivo Optical Imaging
Published on: January 5, 2015
Portable Near-Infrared Fluorometer for a Liposomal Blood Lactate Assay
Natalie Guirguis1, Arturo Israel Machuca-Parra1, Simon Matoori1
1Faculté de Pharmacie, Université de Montréal, 2940 Chemin de Polytechnique, Montreal, Quebec H3T 1J4, Canada.
Abstract:
In sepsis, plasma lactate is a key biomarker of disease severity, prognosis, and treatment success. However, the median time to result for clinical lactate tests is 3 h. We recently reported a near-infrared fluorescent (NIRF) blood lactate assay that relies on a two-step enzymatic reaction in a liposomal reaction compartment. This assay was optimized in human blood and was capable of quantifying lactate in fresh capillary blood from human volunteers at clinically relevant concentrations in 2 min. However, these studies were performed with a tabletop fluorescence plate reader. For translation to the point of care, the liposomal lactate assay needs to be combined with a small portable NIR fluorometer. Portable NIR fluorometers were successfully used for the analysis of skin and soil samples, but reports for blood metabolite assays are scarce. We aimed at testing the performance of the liposomal lactate assay in combination with a commercial small portable NIR fluorometer. First, we tested the fluorophore of the liposomal lactate assay using the NIR dye sulfo-cyanine 7; we observed strong fluorescence signals and high linearity. Second, we performed the liposomal lactate assay in lactate-spiked human arterial blood using the portable fluorometer as the detector and observed strong and highly linear lactate sensing at clinically relevant lactate concentrations after 2 min. Finally, spiking fresh mouse blood with three clinically relevant lactate concentrations led to a significantly different response to all three concentrations after 5 min. These results highlight the usefulness of the tested portable NIR fluorometer for the liposomal lactate assay and motivate a clinical evaluation of this rapid and easy-to-use lactate assay.