Evaluation of the analytical sensitivity of ACON and LumiraDx SARS-CoV-2 rapid antigen tests using samples with
Brad S Karon1, Leslie J Donato1, Ann M Moyer1
1Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester MN 55905 USA.
Background:
Analytical sensitivity of 2 rapid antigen tests was evaluated for detection of presumed SARS-CoV-2 Omicron variants and earlier variants of concern.
Methods:
A total of 152 SARS-CoV-2 RNA positive samples (N and ORF1ab positive but S gene negative) were tested for SARS-CoV-2 antigen by ACON lateral flow and LumiraDx fluorescence immunoassays. Sensitivity within 3 viral load ranges was compared among these 152 samples and 194 similarly characterized samples collected prior to the circulation of the Delta variant (pre-Delta).
Results:
Antigen was detected in >95% of pre-Delta and presumed Omicron samples for both tests at viral loads >500,000 copies/mL, and 65 to 85% of samples with 50,000-500,000 copies/mL. At viral load <50,000 copies/mL, antigen tests showed better sensitivity in detecting pre-Delta compared to Omicron variants. LumiraDx was more sensitive than ACON at low viral load.
Conclusions:
Antigen tests had decreased sensitivity for detecting presumed Omicron compared to pre-Delta variants at low viral load.