In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
Validation of a clinically applicable device for fast and accurate quantification of blood volume
Andreas Breenfeldt Andersen1, Thomas C Bonne2, Joar Hansen3
1Department of Public Health, Research Unit for Exercise Biology, Aarhus University, Aarhus, Denmark.
Introduction:
Determination of blood volume (BV) using the dual-isotope (e.g., 99m Tc-labeled red blood cells [99m Tc-RBC] and 125 I-labeled human serum albumin [125 I-HSA]) injection method is limited in medicine due to the long isotope half-life. However, BV has been determined in laboratory settings for 100 years using the carbon monoxide (CO)-rebreathing-based procedure, which allows frequent BV measurements.
Methods:
We investigated the reliability and accuracy of a semi-automated CO-rebreathing device by comparing it against the dual-isotope methodology and its ability to detect a known blood removal. In study A, BV was determined three times in ~2 h; twice using the device with rebreathing protocols lasting 2 (CO2min ) and 10 min (CO10min ) and once with the dual-isotope technique. In study B, the accuracy of the device was assessed by its ability to detect a 2% removal of BV.
Results:
A good correlation was observed between both the CO-rebreathing protocols (r2 = 0.89-0.98; p < 0.001) and the dual-isotope approach (r2 = 0.89-0.95; p < 0.001). In absolute terms BV was 425 ± 263 mL and 491 ± 388 mL lower (p < 0.001) when quantified with the dual-isotope compared to the CO-rebreathing protocols. When reducing BV by 132 ± 25 mL (2%), the device quantified a lower (p < 0.001) BV of 150 ± 45 mL.
Conclusion:
This study emphasizes that the semi-automated device accurately determines small changes (i.e., 2%) in BV and that a high correlation with the dual-isotope methodology exists. The findings are clinically relevant owing to the method's simple and fast nature (the absence of radioactive tracers and reduced time requirements, i.e., ~15 min vs. ~180 min) and the possibility for repeated measurements within a single day.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...