Comparison of insertion characteristics of LMA ProSeal and Ambu AuraGain in adult patients under controlled
Rashmi Salhotra1, Anurag Padhy1, Rajesh S Rautela1
1Department of Anaesthesiology, Critical Care and Pain Medicine, University College of Medical Sciences and Guru Teg Bahadur Hospital, Delhi, India.
Background And Aims:
Supraglottic devices are preformed or flexible, and the insertion characteristics of the two types may be different. This study aims to compare the insertion characteristics of Ambu AuraGain (AAG), preformed) and LMA ProSeal (PLMA), flexible, requires an introducer tool for placement.
Methods:
Forty American Society of Anesthesiologists (ASA) physical status I/II patients of either sex between 18 and 60 years with no anticipated airway difficulty were randomly allocated to either group AAG or PLMA (n = 20 each). Pregnant females, known case of chronic respiratory disorders and gastroesophageal reflux were excluded. After induction of anaesthesia and muscle relaxation, appropriately sized AAG or PLMA was inserted. Time for successful insertion (primary outcome), ease of device insertion and gastric drain insertion, first attempt success rate (secondary outcomes) were recorded. Statistical analysis was done using SPSS version 20.0. Quantitative parameters were compared using Student's t-test, and qualitative parameters were compared using Chi-square test. A P value of <0.05 was considered significant.
Results:
Time taken for successful insertion of PLMA was 22.94 ± 6.12 s and for AAG was 24.32 ± 4.96 s, (P = 0.458). Device insertion was significantly easy in PLMA group (P < 0.002). First attempt success rate was achieved in 17 (94.4%) cases in PLMA group compared to 15 (78.9%) cases in AAG group (P = 0.168). Ease of drain tube insertion was comparable among the groups (P = 0.298). The haemodynamic variables were also comparable.
Conclusion:
PLMA is easier to insert as compared to AAG, but the insertion time and first attempt success rate are similar. The preformed curvature in AAG does not provide any added advantage over the non-preformed PLMA.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...