Visualization of the distortion induced by nonlinear noise reduction in computed tomography
Joel Larsson1,2, Magnus Båth1,3, Anne Thilander-Klang1,3
1University of Gothenburg, Sahlgrenska Academy, Institute of Clinical Sciences, Department of Medical Radiation Sciences, Gothenburg, Sweden.
Purpose:
We developed a method to visualize the image distortion induced by nonlinear noise reduction algorithms in computed tomography (CT) systems.
Approach:
Nonlinear distortion was defined as the induced residual when testing a reconstruction algorithm by the criteria for a linear system. Two types of images were developed: a nonlinear distortion of an object () image and a nonlinear distortion of noise () image to visualize the nonlinear distortion induced by an algorithm. Calculation of the images requires access to the sinogram data, which is seldomly fully provided. Hence, an approximation of the image was estimated. Using simulated CT acquisitions, four noise levels were added onto forward projected sinograms of a typical CT image; these were noise reduced using a median filter with the simultaneous iterative reconstruction technique or a total variation filter with the conjugate gradient least-squares algorithm. The linear reconstruction technique filtered back-projection was also analyzed for comparison.
Results:
Structures in the image indicated contrast and resolution reduction of the nonlinear denoising. Although the approximated image represented the original image well, it had a higher random uncertainty. The image for the median filter indicated both stochastic variations and structures reminding of the object while for the total variation filter only stochastic variations were indicated.
Conclusions:
The developed images visualize nonlinear distortions of denoising algorithms. The object may be distorted by the noise and vice versa. Analyzing the distortion correlated to the object is more critical than analyzing a distortion of stochastic variations. The absence of nonlinear distortion may measure the robustness of the denoising algorithm.
More Related Videos
13:44Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...