Magnetic resonance imaging is safe in patients with left bundle branch pacing
Breno Bernardes-Souza1, Raj V Bhatt1, Aileen Joy S Bayoneta2
1UCLA Cardiac Arrhythmia Center, David Geffen School of Medicine at UCLA, Los Angeles, California, USA.
Background:
The use of left bundle branch pacing (LBBP) has dramatically increased since it was first described in 2016, but to date there are no published data on the safety of performing magnetic resonance imaging (MRI) in these patients.
Methods:
Patients with LBBP who underwent MRI between January 2016 and October 2022 were retrospectively studied in our clinical center, which has a special program for imaging patients with cardiac devices. All patients underwent close cardiac monitoring throughout the MRI scans. Occurrence of arrhythmias or other adverse effects during MRI were assessed. LBBP lead parameters immediately pre- and post-MRI and at an outpatient follow-up were compared.
Results:
Fifteen patients with LBBP underwent a total of 19 MRI sessions during the study period. Lead parameters did not significantly change after the MRI or on follow-up, which took place at a median of 91 days after the MRI. No patient developed arrhythmias during the MRI sessions, and no adverse effects such as lead dislodgement were reported.
Conclusion:
Although larger studies are necessary to verify our findings, MRI in patients with LBBP appears safe based on this initial case series.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System V: CT