Related Experiment Video
Updated: Jul 26, 2025

Ischemia-reperfusion Model of Acute Kidney Injury and Post Injury Fibrosis in Mice
Published on: August 9, 2013
Artificially raised creatinine concentrations due to analytical interference for samples contaminated with total
Jinny Jeffery1, Helen Millar1, Rachel Marrington2
1Derriford Combined Laboratory, Derriford Hospital, Plymouth, UK.
Background:
A sample received in the laboratory from a patient receiving total parenteral nutrition (TPN) indicated that the patient may have renal dysfunction, but the results were not considered to be reliable enough to report. Investigations using a reference method for measurement of creatinine confirmed positive interference in the creatinine assay and distribution of samples via an External Quality Assessment (EQA) Scheme showed that this positive interference was method dependent.
Methods:
Residual TPN fluid (Nutriflex Lipid Special) left in the bag after the patient had completed the infusion was collected and added to a patient serum pool in increasing amounts and distributed to different laboratories for analysis of creatinine and glucose through an EQA Scheme.
Results:
Positive interference in a number of different creatinine assays was identified as a result of a component in the TPN fluid. Positive interference from high concentrations of glucose has been demonstrated to be a cause for falsely high results in Jaffe creatinine assays.
Conclusions:
The concern would be that a sample contaminated with TPN fluid would have both abnormal electrolytes and creatinine concentrations and give the impression that the patient was in renal failure due to analytical interference in the creatinine assay and laboratory staff need to be aware of this problem.
More Related Videos
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
11:17Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
Related Concept Videos
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Acute Kidney Injury V: Interprofessional Care
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...