Investigation of Computed Tomography Numbers on Multiple Imaging Systems using Single and Multislice Methods
Akash Mehta1, Daniel Arrington1, Prabhakar Ramachandran1
1Department of Radiation Oncology, Cancer Services, Princess Alexandra Hospital, Woolloongabba, Queensland, Australia.
This study found minimal Hounsfield unit (HU) variation with fan-beam computed tomography (FCT), suggesting single-slice methods suffice for calibration curves. However, cone-beam CT (CBCT) systems, especially Gamma Knife, show significant HU variation, impacting dose calculations.
Area of Science:
- Medical Physics
- Radiological Imaging
Background:
- Accurate Hounsfield Unit (HU) values are crucial for CT-based electron density calculations in radiotherapy.
- Variations in HU can arise from different imaging techniques and acquisition methods, potentially affecting treatment planning.
Purpose of the Study:
- To evaluate Hounsfield value variations using single and multi-slice methods across fan-beam CT (FCT), linac-based cone-beam CT (CBCT), and Icon-CBCT.
- To assess the impact of these variations on dose calculations.
Main Methods:
- Scanned Gammex and Advanced Electron Density (AED) phantoms using Toshiba CT, five linac-based CBCT systems, and Leksell Gamma Knife Icon.
- Developed in-house MATLAB software to analyze HU statistics and longitudinal trends.
- Assessed dosimetric changes using a CIRS Model 605 Radiosurgery Head Phantom.
Main Results:
- FCT showed minimal HU variation (± 3 HU) along the long axis, consistent across clinical protocols.
- Linac-based CBCTs exhibited insignificant variation among themselves, but notable HU deviations were observed in specific cases (e.g., water insert).
- Gamma Knife CBCT demonstrated the most significant HU variation along the long axis, with dose differences up to 1 Gy compared to CT.
Conclusions:
- Single-slice HU calibration is sufficient for FCT due to minimal variation.
- CBCT systems, particularly Gamma Knife, show significant HU variations that can impact dose calculations.
- Multiple slice assessment of Hounsfield values is recommended for CBCT-based dose calculations.
More Related Videos
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
08:46Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT