Related Experiment Video
Updated: Jul 25, 2025

Quantification of Oculomotor Responses and Accommodation Through Instrumentation and Analysis Toolboxes
Published on: March 3, 2023
Do ophthalmology residents know how to check the calibration of a Goldmann applanation tonometer?
Aaron Carlisle1, Sofia Habib2, Zhiheng Lin2
1Norwich Medical School, University of East Anglia, Norwich, UK. aaron.carlisle@doctors.org.uk.
Purpose:
The purpose of this observational study was to determine whether ophthalmology residents know how to check Goldmann applanation tonometer (GAT) calibration.
Methods:
The step-by-step technique for checking the calibration of a GAT was taken from the manufacturer's manual and developed into a mark sheet. Ophthalmology residents in years 2-8 of training from 11 hospitals were individually observed and assessed checking calibration of a GAT. Participation was voluntary. Contact between participants was minimised to prevent communication about the study.
Results:
Sixty-eight per cent (n = 30) of eligible ophthalmology residents (years 2-8) from 11 hospitals (three teaching hospitals and eight local general hospitals) were observed checking GAT calibration. Only 33% (n = 10; 95% CI: 16-50%) of ophthalmology residents were able to correctly check GAT calibration. Those participants who were previously taught (p = 0.046) or assessed (p = 0.015) were more likely to be successful in GAT calibration.
Conclusions:
Most ophthalmology residents were unable to correctly check GAT calibration. Although better than previously published results, this observational study shows that further training and assessment is required for ophthalmology residents to learn the technique of checking GAT calibration.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure