Related Experiment Video
Updated: Jul 25, 2025

A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
Deep learning-based monitoring technique for real-time intravenous medication bag status
Young Jun Hwang1, Gun Ho Kim2, Min Jae Kim1
1Department of Biomedical Engineering, School of Medicine, Pusan National University, Yangsan, Korea.
Abstract:
Accidents related to the administration of intravenous (IV) medication, such as drug overdose/underdose, drug/patient mis-identification, and delayed bag exchange, occur consistently in clinical fields. Several previous studies have suggested various contact-sensing and image-processing methodologies; however, most of them can increase the workload of nursing staffs during the long-term, continuous monitoring. In this study, we proposed a smart IV pole that can monitor the infusion status of up to four IV medications (patient/drug identification, and liquid residue) with various sizes and hanging positions to reduce IV-related accidents and improve patient safety with the least additional workload; the system consists of 12 cameras, one code scanner, and four controllers. Two types of deep learning models for automated camera selection (CNN-1) and liquid residue monitoring (CNN-2), and three drug residue estimation equations were implemented. The experimental results demonstrated that the accuracy of identification code-checking (60 tests) was 100%. The classification accuracy and the mean inference time of CNN-1 (1200 tests) were 100% and 140 ms. The mean average precision and the mean inference time of CNN-2 (300 tests) were 0.94 and 144 ms. The average error rates between the alarm setting (20, 30, and 40 mL) and the actual drug residue when the alarm first generated were 4.00%, 7.33%, and 4.50% for a 1,000 mL bag; 6.00%, 4.67%, and 2.50% for a 500 mL bag; and 3.00%, 6.00%, and 3.50% for a 100 mL bag, respectively. Our results suggest that the implemented AI-based prototype IV pole is a potential tool for reducing IV-related accidents and improving in-hospital patient safety.
Supplementary Information:
The online version contains supplementary material available at 10.1007/s13534-023-00292-w.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Errors occurring during blood pressure monitoring
Several factors...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...