Estimation of change limits (deltacheck) in clinical laboratory
Maria-José Castro-Castro1, Lourdes Sánchez-Navarro1
1Clinical Laboratory, Bellvitge University Hospital, Barcelona, Catalonia, Spain.
Objectives:
Change limits, more commonly called delta check, are those in which a change in a patient's measured result in relation to their corresponding preceding measurement is suspected of being erroneous and should be considered as a doubtful result. The aim of this study was to provide change limits for some biochemical and haematological quantities to detect doubtful measured results and to assess its effectiveness to detect erroneous results for their application in and the standardization of the plausibility control.
Methods:
Change limits have been estimated for 13 biochemical and 6 haematological quantities. For each quantity, relative differences (D), expressed as a percentage between the two consecutive measured results from the same patient (from scheduled laboratory requests), were calculated. From these differences (D), the p5 and p95 percentiles of the data distribution were calculated. To assess the effectiveness of the change limits to detect laboratory errors, 43 erroneous laboratory reports, containing different biochemical and haematological quantities, were obtained from the standard laboratory plausibility control procedure.
Results:
From the 43 erroneous laboratory reports, 31 (72%) were due to endovenous administration errors and 12 (28%) were due to mislabeling errors. All erroneous laboratory reports were detected when the change limits of the quantities were combined and applied together.
Conclusions:
The best combination of quantities, which detect all the erroneous reports in the same specimen were: potassium, albumin, creatinine, glucose and haemoglobin.
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Uncertainty in Measurement: Reading Instruments
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Data Validation
Key parameters for method validation include:
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...