Mask ventilation using volume-targeted neonatal ventilator for neonatal resuscitation: a randomised cross-over
Brenda Hiu Yan Law1, Tina Madani Kia2, Faith Trinh2
1Department of Pediatrics, University of Alberta, Edmonton, Alberta, Canada blaw2@ualberta.ca.
This simulation study found that while using a neonatal ventilator for mask ventilation is feasible, it increased healthcare professionals' perceived workload and stress compared to T-piece resuscitators. Human factors require careful consideration before clinical use.
Area of Science:
- Neonatal Resuscitation
- Medical Simulation
- Human Factors Engineering
Background:
- Mask ventilation is a critical component of neonatal resuscitation.
- Traditional T-piece resuscitators are commonly used, but novel methods are being explored.
- Understanding human factors is essential for safe and effective implementation of new medical devices.
Purpose of the Study:
- To compare a novel ventilator-based mask ventilation method with standard T-piece ventilation using simulations.
- To evaluate human factors, including workload and visual attention, associated with each method.
- To inform the clinical application of ventilator-based mask ventilation in neonates.
Main Methods:
- A randomized cross-over simulation study was conducted with experienced healthcare professionals (HCPs).
- Participants performed simulated preterm neonatal resuscitations using either ventilator-based volume-targeted positive pressure mask ventilation (VTV-PPV) or T-piece positive pressure ventilation (T-piece PPV).
- Subjective workload (SURG-TLX) and visual attention (VA) were measured using eye-tracking glasses.
Main Results:
- HCPs reported significantly higher workload scores for VTV-PPV, particularly in mental demand, physical demand, task complexity, and situational stress.
- Visual attention distribution was similar between the two methods, despite longer simulation times for VTV-PPV.
- While VTV-PPV was rated positively overall, participants noted challenges with ergonomics, cognition, and teamwork.
Conclusions:
- Ventilator-based volume-targeted positive pressure mask ventilation is a feasible approach for neonatal resuscitation.
- Implementation requires careful consideration of human factors, including workload and potential ergonomic and cognitive challenges.
- Further research and refinement are needed to optimize this novel ventilation method for clinical practice.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...