Artifacts in two-dimensional shear wave elastography of liver.
Hui-Peng Wang1, Peng-Chao Zheng1, Xue-Mei Wang1
1Department of Ultrasound, The First Hospital of China Medical University, Shenyang 110001, Liaoning Province, China.
World Journal of Gastroenterology
|June 28, 2023
Summary
Artifacts are common in two-dimensional shear wave elastography (2-D SWE) for liver stiffness (LS) measurement, especially with novice operators. These artifacts can overestimate LS and decrease measurement reliability.
Area of Science:
- Medical imaging
- Hepatology
- Diagnostic techniques
Background:
- Two-dimensional shear wave elastography (2-D SWE) is widely used for liver stiffness (LS) assessment.
- Artifacts are frequently encountered in 2-D SWE but are often unrecognized.
- The impact of these artifacts on LS measurements requires further investigation.
Purpose of the Study:
- To investigate the prevalence and influence of artifacts during 2-D SWE liver examinations.
- To compare artifact occurrence between novice and expert operators.
- To evaluate the effect of artifacts on the accuracy and reliability of LS measurements.
Main Methods:
- 158 patients with chronic liver disease underwent 2-D SWE by a novice and an expert.
- Elastograms were analyzed for artifact presence and location (top-left, top-right, bottom-left, bottom-right).
- Liver stiffness values (LSVs) and stability were compared between elastograms with most artifacts (EMA) and least artifacts (ELA).
Main Results:
- Artifacts were significantly more frequent in novice (51.7%) versus expert (19.6%) operators.
- The bottom-left quadrant showed the highest artifact frequency for both operators.
- EMAs exhibited higher LSVs and standard deviations, and lower stability indices compared to ELAs, impacting reliability.
Conclusions:
- Artifacts are a significant issue in 2-D SWE for LS measurement, particularly for less experienced operators.
- The presence of artifacts can lead to overestimation of liver stiffness.
- Artifacts reduce the repeatability and reliability of LS measurements obtained via 2-D SWE.