Commissural and coronary alignment of the Venus-A valve after transcatheter aortic valve replacement: a retrospective
Siyu Xi1, Haishan Zhang1, Yuan Gao1
1Department of Cardiology, The First Hospital of China Medical University, Shenyang, China.
Insights
Transcatheter aortic valve replacement (TAVR) with the Venus-A valve using a standard technique did not achieve commissural or coronary alignment. Further methods are needed to optimize Venus-A valve positioning for improved patient outcomes.
Area of Science:
- Cardiovascular Medicine
- Interventional Cardiology
- Medical Imaging
Background:
- Neo-commissural orientation of transcatheter heart valves (THV) impacts coronary obstruction, THV durability, and future coronary access post-TAVR.
- Specific orientations for Evolut R/Pro and Acurate Neo valves are known to improve commissural alignment.
- Methods for achieving commissural alignment with the Venus-A valve are not established.
Purpose of the Study:
- To evaluate commissural and coronary alignment of the Venus-A self-expanding valve after TAVR.
- To assess alignment using a standard system delivery technique.
Main Methods:
- Retrospective cross-sectional study of 45 TAVR patients.
- Analysis of pre- and post-procedural CT scans.
- Categorization of commissural misalignment (CMA) and coronary overlap (CO) based on defined angle deviations.
Main Results:
- Significant commissural misalignment observed: 20.0% aligned, 33.3% mild CMA, 26.7% moderate CMA, 20.0% severe CMA.
- High incidence of coronary overlap: 24.4% severe CO with left main, 28.9% with right coronary artery.
- 46.7% of patients experienced coronary overlap with one or both coronary arteries.
Conclusions:
- Standard delivery technique for Venus-A valve does not achieve commissural or coronary alignment.
- New strategies are required to optimize Venus-A valve positioning and alignment.
Background:
Previous studies have shown that neo-commissural orientation of transcatheter heart valve (THV) can influence coronary obstruction during transcatheter aortic valve replacement (TAVR), long-term durability of THV, and coronary artery access for reintervention after TAVR. Specific initial orientations of Evolut R/Pro and Acurate Neo aortic valves can improve commissural alignment. However, the method of achieving commissural alignment with the Venus-A valve remains unknown. Therefore, this study aimed to evaluate the extent of commissural and coronary alignment of the Venus-A self-expanding valve after TAVR using a standard system delivery technique.
Methods:
A retrospective cross-sectional study was performed. At the time of enrollment, patients who underwent pre- and post-procedural electrocardiographically-gated contrast-enhanced CT with a second-generation 64-row multidetector scanner were selected for the study. Commissural alignment was categorized as aligned (0-15° angle deviation), mild (15-30°), moderate (30-45°), or severe (45-60°) commissural misalignment (CMA). Coronary alignment was categorized as having no coronary overlap (CO) (>35°), moderate CO (20-35°), or severe CO (≤20°). The results were represented as proportions to assess the extent of commissural and coronary alignment.
Results:
Forty-five TAVR patients were ultimately included in the analysis. THVs were shown to be randomly implanted: 20.0% of THVs were aligned, 33.3% had mild CMA, 26.7% had moderate CMA, and 20.0% had severe CMA. The incidence of severe CO was 24.4% with the left main coronary artery, 28.9% with the right coronary artery, 6.7% with both coronary arteries, and 46.7% with one or both coronary arteries.
Conclusions:
The results showed that commissural or coronary alignment could not be achieved with the Venus-A valve using a standard system delivery technique. Therefore, specific methods to attain alignment with the Venus-A valve need to be identified.