Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Accuracy of wearable electronic device compared to manual and automatic methods of blood pressure determination
Lindercy Francisco Tomé de Souza Lins1, Ellany Gurgel Cosme do Nascimento1, José Antonio da Silva Júnior2
1Faculdade de Ciências da Saúde, Universidade Do Estado Do Rio Grande Do Norte, Atirador Miguel Antonio da Silva St, Mossoró, RN, 59607-360, Brazil.
Abstract:
Blood pressure (BP) is the main biomarker for monitoring patients, as its lack of control above values considered normal is a modifiable risk factor for target organ damage. The aim of this study is to evaluate the accuracy of the wearable electronic device photoplethysmography technology (PPG) Samsung Galaxy Watch 4 in determining BP in young patients compared to manual and automatic methods of BP determination. This is a quantitative and cross-sectional study, following validation protocols for wearable devices and BP measurement. It was carried out with twenty healthy young adults, in which BP was measured using four instruments, namely, standard sphygmomanometer device (manual), automatic arm oscillometric device (reference), wrist oscillometric device, and Smartwatch PPG. Eighty systolic blood pressure (SBP) and diastolic blood pressure (DBP) readings were observed. SBP means manual 118 ± 2.20,arm 113 ± 2.54, wrist 118 ± 2.51, and PPG (smartwatch) 113 ± 2.58. Among means, arm and PPG difference is 0.15, arm and wrist 4.95, arm and manual 4.45 wrist with PPG. The mean DBP manual 76.7 ± 1.84, arm 73.6 ± 1.92, wrist 79.3 ± 1.87, and PPG 72.2 ± 1.38. Among means, the difference between the arm and PPG is 1.4 and arm and hand 3.5 mmHg. The correlation shows PPG with manual, arm, and wrist. There was a strong SBP correlation and a moderate DBP correlation between the methods tested, evidencing the accuracy of the PPG smartwatch in relation to the reference method.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement