Accelerating TOF-MRA: The impact of the combined use of compressed sensitivity encoding and spiral imaging
Kosuke Morita1, Hiroyuki Uetani2, Takeshi Nakaura2
1Department of Radiology, Kumamoto University Hospital, Honjo 1-1-1, Kumamoto, Japan.
Purpose:
To evaluate the image quality of the combined technique of compressed sensitivity encoding (CS) and spiral imaging in time-of-flight magnetic resonance angiography (TOF-MRA), which is approximately 2.5 times faster than conventional methods.
Methods:
Twenty volunteers underwent four TOF-MRA sequences: sensitivity encoding (SENSE) with acceleration factor of 4 (acquisition time: 4:55 min), CS with acceleration factor of 10.9, and spiral and CS-spiral (both 1:55 min). A quantitative image analysis (signal-to-noise ratio [SNR], contrast, and full width at half maximum [FWHM] edge criterion measurements) was performed on four TOF sequences. For qualitative image analysis, two board-certified radiologists evaluated the overall depiction of the proximal, intermediate, and distal branches in CS, spiral, and CS-spiral images using SENSE as a reference.
Results:
The SNR of BA in spiral and CS-spiral imaging was significantly lower than that in SENSE (p = 0.009). The contrasts of ACA and BA in CS-spiral were significantly higher and those in spiral were significantly lower than those in SENSE (p < 0.001). The FWHM in the CS image was significantly higher than that of SENSE; however, no significant differences were observed between the spiral or CS-spiral and SENSE. In qualitative analysis, the depiction of proximal vascular branches was significantly impaired in spiral than in others and that of distal vascular branches was significantly impaired in CS than in others (p < 0.001).
Conclusions:
In TOF-MRA, which is approximately 2.5 times faster than conventional methods, the combined use of CS and spiral imaging demonstrated an improvement in image quality compared to either CS or spiral imaging alone.
Summary Statement:
The image quality of Compressed SENSE and spiral imaging is particularly poor in the proximal and distal vascular branches, respectively at an extremely high acceleration factor; however, CS-spiral provided stable image quality in all regions as compared with the SENSE technique.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...