Related Experiment Video
Updated: Jul 24, 2025

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Evaluation of a planar diode matrix for SRS patient-specific QA in comparison with GAFchromic films
Erminia Infusino1, Anna Ianiro1, Stefano Luppino1
1Department of Medical Physics, IRCCS Regina Elena National Cancer Institute - Rome, Italy.
Purpose:
We validate the routine use of a two-dimensional (2D) diode matrix for patient specific pre-treatment verification for Cyberknife (CK) stereotactic radiosurgery and to compare it with film dosimetry.
Materials And Method:
A total of 46 patients were selected according to the most frequent diseases treated at our institution with the CK system, that is, brain metastases, meningiomas, spine metastases, and prostate tumors. All cases were evaluated with GAFChromic EBT-3 films and SRS MapCHECK for Fixed cone, IRIS, and MLC collimators of the CK.
Results:
The highest mean passing rate was observed for the SRS MapCHECK system compared to films. In order to assess if the two techniques provide statistically different results, a Wilcoxon Signed-Rank non-parametric test was performed (p < 0.05) and we found gamma values significantly lower for EBT-3 films with respect to the SRS MapCHECK. We noticed a moderately significant association between the two techniques using Spearman's rank correlation coefficient (rs > 0.4). We also performed the Bland-Altman statistical method: less than 5% of the differences resulted outside the range (mean ± 1.96 × SD), so the two methods can be considered interchangeable within the combined inaccuracy.
Conclusions:
The use of SRS MapCHECK for CK patient specific quality assurance (QA) is feasible for a variety of clinical districts and could be reliably used as a replacement for radiochromic films.