Quantification of Blood Viscoelasticity under Microcapillary Blood Flow
1Department of Mechanical Engineering, Chosun University, 309 Pilmun-daero, Dong-gu, Gwangju 61452, Republic of Korea.
Micromachines
|July 8, 2023
Summary
This study introduces a novel method to measure blood viscoelasticity by separating sample compliance from microfluidic system effects. The new approach accurately quantifies red blood cell elasticity, improving blood flow analysis.
Area of Science:
- Biomedical Engineering
- Microfluidics
- Rheology
Background:
- Quantifying blood elasticity is crucial for understanding pulsatile blood flow.
- Existing single-compliance models are significantly influenced by microfluidic system properties (e.g., soft channels, flexible tubing).
- This limitation hinders accurate assessment of intrinsic blood viscoelasticity.
Purpose of the Study:
- To develop a novel method for disentangling blood viscoelasticity from microfluidic system compliance.
- To introduce two distinct compliance coefficients: one for the microfluidic system (C1) and one for red blood cell (RBC) elasticity (C2).
- To enable more accurate measurements of blood viscoelasticity independent of the measurement device.
Main Methods:
- Utilized a coflowing microfluidic channel for blood viscoelasticity estimation.
- Derived a governing equation for the interface using fluidic circuit modeling and solved it analytically.
- Employed nonlinear curve fitting to determine the two compliance coefficients (C1 and C2).
Main Results:
- The ratio of RBC elasticity to microfluidic system compliance (C2/C1) was estimated between 10.9 and 20.4, varying with channel depth.
- Polydimethylsiloxane (PDMS) channel depth increased both compliance coefficients; outlet tubing decreased C1.
- Both compliance coefficients and blood viscosity showed significant variations with homogeneous versus heterogeneous hardened RBCs.
Conclusions:
- The proposed two-compliance coefficient method effectively distinguishes blood viscoelasticity from microfluidic system effects.
- This technique can accurately detect changes in blood properties and microfluidic systems.
- Future applications include the detection of red blood cell subpopulations in patient blood samples.
Related Concept Videos
Blood Flow
70.3K
Blood is pumped by the heart into the aorta, the largest artery in the body, and then into increasingly smaller arteries, arterioles, and capillaries. The velocity of blood flow decreases with increased cross-sectional blood vessel area. As blood returns to the heart through venules and veins, its velocity increases. The movement of blood is encouraged by smooth muscle in the vessel walls, the movement of skeletal muscle surrounding the vessels, and one-way valves that prevent backflow.
70.3K
Viscosity of Fluid
475
Viscosity measures the resistance a fluid offers to flow and deformation. It results from internal friction between layers of fluid moving relative to one another. Dynamic viscosity, denoted by the Greek letter mu (μ), quantifies the force needed to move one fluid layer over another. For Newtonian fluids like water and air, the relationship between the shearing stress and the rate of shearing strain is linear, meaning their viscosity remains constant regardless of the applied stress.