Related Experiment Video
Updated: Jul 24, 2025

Bidirectional Electrical and Optoelectronic Interfaces in Healthy and Ischemic Ex Vivo Rat Hearts
Published on: July 18, 2025
Rapid cardiac ischemia detection with an epicardial graphene probe
Grzegorz Suwalski1, Marek Galanty2, Beata Degórska2
1Department of Cardiac Surgery, Military Institute of Medicine, Warsaw, Poland.
Introduction:
In this study, a new probe was designed to enable electrocardiography of a rotated heart during cardiac surgery when skin electrodes became non-functional. This probe adhered non-invasively to the epicardium and collected the ECG signal independently from the position of the heart. The study compared the accuracy of cardiac ischemia detection between classic skin and epicardial electrodes in an animal model.
Methods:
Using six pigs, an open chest model was devised with cardiac ischemia induction by coronary artery ligation in two non-physiologic heart positions. Both the accuracy and the time of detection of electrocardiographic symptoms of acute cardiac ischemia were compared between skin and epicardial methods of signal collection.
Results:
Heart rotation to expose either the anterior or the posterior wall resulted in a distortion or loss of the ECG signal collected by skin electrodes after coronary artery ligation, standard skin ECG monitoring did not reveal any ischemia symptoms. Attachment of an epicardial probe on the anterior and posterior walls helped in the recovery of the normal ECG wave. After ligation of the coronary artery, the epicardial probes recorded cardiac ischemia within 40 s.
Discussion:
This study highlighted the effectiveness of ECG monitoring with epicardial probes in a rotated heart. It can be concluded that epicardial probes can detect the presence of acute ischemia of a rotated heart when skin ECG monitoring becomes ineffective.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Acute Coronary Syndrome III: Diagnostic Studies