Optimal Abdominal CT Image Quality in Non-Lean Patients: Customization of CM Injection Protocols and Low-Energy
Francesco Macri1, Elina Khasanova1, Bonnie T Niu2
1Department of Radiology, Geneva University Hospitals, University of Geneva, 1211 Geneva, Switzerland.
Customized contrast media injection protocols significantly reduce contrast volume and injection rate in abdominopelvic CT scans. This approach enhances image quality for both single-energy CT (SECT) and dual-energy CT (DECT) compared to standard fixed protocols.
Area of Science:
- Radiology
- Medical Imaging
- Computed Tomography
Background:
- Standard abdominopelvic CT scans often use fixed contrast media (CM) volumes.
- Optimizing CM administration is crucial for balancing image quality and patient dose.
Purpose of the Study:
- To compare the image quality of abdominopelvic single-energy CT (SECT) and dual-energy CT (DECT) using customized CM injection protocols versus standard fixed protocols.
Main Methods:
- Retrospective analysis of 91 patients (SECT-100 kVp) and 83 patients (DECT-65 keV) with customized injection protocols.
- Comparison with prior abdominopelvic SECT-120 kVp scans using fixed CM injection.
- Customized protocols based on patient weight, contrast concentration, and injection rate formula.
Main Results:
- Customized injection significantly reduced CM volume (9-12 mL) and injection rate (0.2-0.4 mL/s) compared to SECT-120 kVp (p ≤ 0.001).
- Improved attenuation, noise, signal-to-noise, and contrast-to-noise ratios were observed for SECT-100 kVp and DECT-65 keV.
- Qualitative image quality scores were high (≥4) in 98.8% of customized injection cases.
Conclusions:
- Customized CM injection protocols offer a substantial reduction in iodine dose.
- These protocols yield superior image quality in abdominopelvic SECT-100 kVp and DECT-65 keV examinations.
- Weight-based customized protocols represent an effective strategy for optimizing CM administration in CT imaging.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging