Fabrication of Thin Film Silver/Silver Chloride Electrodes with Finely Controlled Single Layer Silver Chloride
Published on: July 1, 2020
Multielectrode Arrays at Wafer-Level for Miniaturized Sensors Applications: Electrochemical Growth of Ag/AgCl
Haosheng Wu1, Robert Krause1, Eshanee Gogoi1
1Fraunhofer Institute for Photonic Microsystems (IPMS), Center Nanoelectronic Technologies (CNT), An der Bartlake 5, 01109 Dresden, Germany.
Abstract:
In this study, a range of miniaturized Ag/AgCl reference electrodes with various layouts were successfully fabricated on wafer-level silicon-based substrates with metallic intermediate layers by precisely controlling the electrochemical deposition of Ag, followed by electrochemical chlorination of the deposited Ag layer. The structure, as well as the chemical composition of the electrode, were characterized with SEM & EDS. The results showed that the chlorination is very sensitive to the applied electric field and background solution. Potentiostatic chlorination, in combination with an adjusted mushroom-shaped Ag sealing deposition, enabled the formation of electrochemical usable Ag/AgCl layers. The stability of the electrodes was tested using open circuit potential (OCP) measurement. The results showed that the reference electrodes stayed stable for 300 s under 3 M KCl solution. The first stage study showed that the stability of the Ag/AgCl reference electrode in a chip highly depends on chip size design, chlorination conditions, and a further protection layer.
More Related Videos
08:03Multi-analyte Biochip MAB Based on All-solid-state Ion-selective Electrodes ASSISE for Physiological Research
Published on: April 18, 2013
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017