Related Experiment Video
Updated: Jul 23, 2025

09:30
Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
19.6K
Simultaneous multislice EPI prospective motion correction by real-time receiver phase correction and coil sensitivity
Bo Li1, Ningzhi Li2, Ze Wang1
1Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland, Baltimore, Maryland, USA.
Magnetic Resonance in Medicine
|July 14, 2023
Summary
Prospective motion correction (PMC) in simultaneous multislice (SMS) echo-planar imaging (EPI) of the brain can be improved by dynamically adjusting receiver phase and resampling coil sensitivities. This reduces motion artifacts in brain scans, enhancing image quality for better clinical insights.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Neuroimaging
- Image Reconstruction
Background:
- Simultaneous multislice (SMS) echo-planar imaging (EPI) is crucial for brain imaging.
- Prospective motion correction (PMC) is essential for mitigating motion artifacts in EPI.
- Existing PMC methods may not fully address artifacts in SMS EPI reconstruction.
Purpose of the Study:
- To enhance image reconstruction for PMC in brain SMS EPI.
- To evaluate the effectiveness of updating receiver phase and resampling coil sensitivities.
Main Methods:
- A camera-based system tracked head motion (3 translations, 3 rotations).
- Dynamic updates of scan position and orientation were performed.
- Receiver phase shifts and resampled coil sensitivity profiles were applied during SMS reconstruction.
Main Results:
- Artifacts in brain SMS EPI scans with PMC were reduced by proposed corrections.
- Correcting coil sensitivity maps improved temporal SNR (tSNR) by 24% for large movements.
- Receiver phase correction boosted tSNR by 50% in a low-motion scan.
Conclusions:
- Dynamically adjusting receiver phase and coil sensitivity profiles minimizes motion artifacts in SMS EPI.
- This method is valuable for improving SMS EPI scans with subject motion.