Radiation dose considerations in digital radiography with an anti-scatter grid: A study using adult and pediatric
Hiroki Kawashima1, Katsuhiro Ichikawa1, Azusa Kitao2
1Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Japan.
Background:
When using an anti-scatter grid, a decrease in receptor dose caused by its X-ray absorption seems to lead to the misperception that radiation dose needs to be increased even in digital radiography (DR).
Objective:
To demonstrate that there is no need to increase radiation dose in DR with a grid, based on a visual evaluation using an adult and a pediatric abdomen phantom (PAD and PPD , respectively).
Materials And Methods:
Phantom images with and without a grid were obtained with exposure parameters determined based on a preliminarily measured signal-to-noise ratio improvement factor (SIF), an index for potential dose reduction when using a grid. In visual evaluation, four radiologists compared phantom images with a grid applied at different dose reduction rates (0% [no reduction], 18%, 36%, and 59% for PAD and 0% and 11% for PPD ) against an image without a grid at the baseline dose (as the reference). They graded the overall image quality of the former relative to that of the latter (reference) on a 3-point scale (3 = better, 2 = almost equal, 1 = worse).
Results:
The mean scores for dose reduction rates of 0%, 18%, 36%, and 59% were 3.00, 3.00, 2.75, and 1.00, respectively, for PAD ; those for 0% and 11% were 2.13 and 1.63, respectively, for PPD . These results support the validity of our view that no dose increase is necessary when using an anti-scatter grid. Actually, there is even a potential for improvement in image quality with dose reduction rates of ≤36% for PAD .
Conclusion:
It is worth reconsidering the necessity of increasing radiation dose in the DR imaging of the adult and pediatric abdomens with an anti-scatter grid.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
X-ray Imaging
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...