Related Experiment Video
Updated: Jul 21, 2025

Transperineal Prostate Biopsy Using a Cone-shaped Double-hole Method with Dual-plane Probe Guidance
Published on: June 6, 2025
A Multicentre Clinical Trial Evaluating a Drop-in Gamma Probe for Minimally Invasive Sentinel Lymph Node Dissection
Wouter Everaerts1, Jochen Walz2, Jose M Abascal Junquera3
1Laboratory of Ion Channel Research, Department of Cellular and Molecular Medicine, KU Leuven, Leuven, Belgium; Department of Urology, University Hospitals Leuven, Leuven, Belgium.
A new drop-in gamma probe achieved a 100% sentinel lymph node detection rate in prostate cancer surgery. This novel device demonstrated improved safety and maneuverability over conventional probes for sentinel lymph node biopsy (SeLNB).
Area of Science:
- Urology
- Surgical Oncology
- Nuclear Medicine
Background:
- Prostate cancer (PCa) management often involves sentinel lymph node biopsy (SeLNB) to assess nodal metastasis.
- Accurate detection of sentinel lymph nodes (SeLNs) is crucial for staging and treatment planning in PCa.
- Existing methods for SeLN detection may have limitations in maneuverability and detection efficacy.
Purpose of the Study:
- To evaluate the safety and performance of a novel drop-in gamma probe for SeLNB in prostate cancer patients.
- To determine the SeLN detection rate using the drop-in gamma probe system.
- To compare the efficacy and safety of the drop-in probe against conventional methods.
Main Methods:
- A prospective, open-label, multicentre clinical trial involving 27 PCa patients undergoing minimally invasive surgery.
- Patients received [99mTc]Tc-nanocolloid injection followed by lymphoscintigraphy.
- Sentinel lymph nodes were identified using the drop-in gamma probe and a rigid laparoscopic gamma probe (RLGP).
Main Results:
- The drop-in gamma probe achieved a 100% SeLN detection rate (95% CI 87.2-100%), identifying 85 SeLNs across 27 patients.
- The device detected two SeLNs missed by the RLGP due to maneuverability limitations.
- No adverse events or complications were associated with the drop-in probe's use.
Conclusions:
- The drop-in gamma probe is a safe and effective tool for SeLNB in prostate cancer.
- The device offers superior maneuverability and enhanced SeLN detection compared to the conventional RLGP.
- This technology meets the performance and safety requirements for clinical SeLNB in PCa.