Comparison of a 2.7-mm and 3.5-mm locking compression plate for ulnar fractures: a biomechanical evaluation
Jenna M Wahbeh1,2, Benjamin V Kelley3, Cyrus Shokoohi2
1The J. Vernon Luck, Sr., M.D. Orthopaedic Research Center, Luskin Orthopaedic Institute for Children in Alliance with UCLA, Los Angeles, CA.
Objectives:
Implant prominence after ulnar fracture fixation may be mitigated by the use of lower profile plates. The biomechanical strength and stability of 2.7-mm and 3.5-mm locking compression plates for fixation were compared.
Methods:
Two fracture conditions, transverse (N = 10) and oblique (N = 10), were evaluated in an in vitro study. Half of the specimens for each condition were fixed with 2.7-mm plates and the other half with 3.5-mm plates, all fixed with conventional dynamic compression mechanisms. Specimens were loaded under ±2 Nm of cyclic axial torsion, then under 10 Nm of cyclic cantilever bending, and bending to failure. Interfragmentary motion and strain were analyzed to determine construct stability as a function of fracture pattern and plate size.
Results:
Interfragmentary motion was significantly larger in all constructs fixed with 2.7-mm plates, compared with 3.5-mm plates (P < 0.01). The 2.7-mm constructs with transverse fractures had the greatest motion, ranging between 5° and 10° under axial rotation and 5.0-6.0 mm under bending. Motions were the lowest for 3.5-mm constructs with oblique fractures, ranging between 3.2 and 4.2 mm under bending and 2°-3.5° for axial rotation. For oblique fractures, the bending moment at ultimate failure was 31.4 ± 3.6 Nm for the 2.7-mm constructs and 10.0 ± 1.9 Nm for 3.5-mm constructs (P < 0.01). Similarly, for transverse fractures, the bending moment was 17.9 ± 4.0 Nm for the 2.7-mm constructs and 9.7 ± 1.3 Nm for the 3.5-mm constructs (P < 0.01).
Conclusions:
Although 3.5-mm plates were more effective at reducing fracture motion, they were consistently associated with refracture at the distal-most screw hole under load to failure. By contrast, 2.7-mm plates plastically deformed despite excessive loads, potentially avoiding a subsequent fracture.
Level Of Evidence:
Level V.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Bones of the Upper Limb: Ulna
Bearings: Problem Solving