Related Experiment Video
Updated: Jul 21, 2025

Rejection of Fluorescence Background in Resonance and Spontaneous Raman Microspectroscopy
Published on: May 18, 2011
Performance assessment of probe-based Raman spectroscopy systems for biomedical analysis
Sean Fitzgerald1,2, Eric Marple3, Anita Mahadevan-Jansen1,2
1Vanderbilt Biophotonics Center, Nashville, TN 37232, USA.
Abstract:
We present a methodology for evaluating the performance of probe-based Raman spectroscopy systems for biomedical analysis. This procedure uses a biological standard sample and data analysis approach to circumvent many of the issues related to accurately measuring and comparing the signal quality of Raman spectra between systems. Dairy milk is selected as the biological standard due to its similarity to tissue spectral properties and because its homogeneity eliminates the dependence of probe orientation on the measured spectrum. A spectral dataset is first collected from milk for each system configuration, followed by a model-based correction step to remove photobleaching artifacts and accurately calculate SNR. Results demonstrate that the proposed strategy, unlike current methods, produces an experimental SNR that agrees with the theoretical value. Four preconfigured imaging spectrographs that share similar manufacturer specifications were compared, showing that their capabilities to detect biological Raman spectra widely differ in terms of throughput and stray light rejection. While the methodology is used to compare spectrographs in this case, it can be adapted for other purposes, such as optimizing the design of a custom-built Raman spectrometer, evaluating inter-probe variability, or examining how altering system subcomponents affects signal quality.
More Related Videos
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...