Purpose:
The main objective of this study was to evaluate the ability of a large field Cadmium Zinc Telluride (CZT) camera to estimate thyroid uptake (TU) on single photon emission computed tomography (SPECT) images with and without attenuation correction (Tomo-AC and Tomo-NoAC) compared with Planar acquisition in a series of 23 consecutive patients. The secondary objective was to determine radiation doses for the tracer administration and for the additional Computed Tomography (CT) scan.
Methods:
Cross-calibration factors were determined using a thyroid phantom, for Planar, Tomo-AC and Tomo-NoAC images. Then Planar and SPECT/CT acquisitions centered on the thyroid were performed on 5 anthropomorphic phantoms with activity ranging from 0.4 to 10 MBq, and 23 patients after administration of 79.2 ± 3.7 MBq of [99mTc]-pertechnetate. We estimated the absolute thyroid activity (AThA) for the anthropomorphic phantoms and the TU for the patients. Radiation dose was also determined using International Commission on Radiological Protection (ICRP) reports and VirtualDoseTMCT software.
Results:
Cross-calibration factors were 66.2 ± 4.9, 60.7 ± 0.7 and 26.5 ± 0.3 counts/(MBq s), respectively, for Planar, Tomo-AC and Tomo-NoAC images. Theoretical and estimated AThA for Planar, Tomo-AC and Tomo-NoAC images were statistically highly correlated (r < 0.99; P < 10-4) and the average of the relative percentage difference between theoretical and estimated AThA were (8.6 ± 17.8), (- 1.3 ± 5.2) and (12.8 ± 5.7) %, respectively. Comparisons between TU based on different pairs of images (Planar vs Tomo-AC, Planar vs Tomo-NoAC and Tomo-AC vs Tomo-NoAC) showed statistically significant correlation (r = 0.972, 0.961 and 0.935, respectively; P < 10-3). Effective and thyroid absorbed doses were, respectively (0.34CT + 0.95NM) mSv, and (3.88CT + 1.74NM) mGy.
Conclusion:
AThA estimation using Planar and SPECT/CT acquisitions on a new generation of CZT large-field cameras is feasible. In addition, TU on SPECT/CT was as accurate as conventional planar acquisition, but the CT induced additional thyroid exposure. Trial registration Name of the registry: Thyroid Uptake Quantification on a New Generation of Gamma Camera (QUANTHYC).
Trial Number:
NCT05049551. Registered September 20, 2021-Retrospectively registered, https://clinicaltrials.gov/ct2/show/record/NCT05049551?cntry=MC&draw=2&rank=4 .
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:55A Practical Guide for the Production and PET/CT Imaging of 68Ga-DOTATATE for Neuroendocrine Tumors in Daily Clinical Practice
Published on: April 17, 2019
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography