An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Do locking plugs improve implant strength? Biomechanical comparison of polyaxial locking constructs with and without
Joni Viitanen1, Robert Quinn2, Matthew Allen3
1Veterinary Specialists Scotland, part of Linnaeus veterinary limited, Livingston, EH54 8AG, United Kingdom. joni.viitanen@gmail.com.
Background:
The objective of this study was to investigate the effects of locking plugs and the biomechanical properties of a 3.5 mm 8-hole polyaxial locking plate in a fracture gap model. Our hypothesis was that locking plugs would increase the strength and stiffness of the construct. Twelve 3.5 mm 8-hole plates were used to evaluate two different construct designs (with locking plugs vs. without locking plugs) with validated bone substitutes in a 25 mm bridging osteosynthesis gap model. Each construct was subjected to a single cycle four-point bending load to failure using a servo-hydraulic testing machine. Bending stiffness, bending strength, and bending structural stiffness were calculated and compared using an unpaired Student´s t-test.
Results:
The plating construct with locking plugs did not show any significant increase in terms of bending stiffness, bending strength, and bending structural stiffness compared to plating construct without locking plugs in a 25 mm gap fracture model during a single cycle four-point bending.
Conclusions:
Under the conditions tested, filling empty plate holes with locking plugs in bridging osteosynthesis does not increase stiffness or strength of the plate-bone construct.