Usefulness of Ag Additional Filter on Image Quality and Radiation Dose for Low-Dose Chest Computed Tomography
Keishin Kawamoto1, Hideyuki Sato, Yosuke Kogure
1From the Department of Radiological Technology, Juntendo University Hospital, Tokyo, Japan.
Objective:
This study aimed to evaluate the effect of a silver (Ag) additional filter on dose characteristics and image quality in low-dose chest computed tomography (CT).
Methods:
A dose evaluation phantom, physical evaluation phantom, and chest phantom were scanned with and without an Ag additional filter. The doses were adjusted so that the displayed the volume CT dose indexes (CTDI vol ) were from 0.3 to 1.6 mGy. For dose characteristics, the spectrum of photon energies and the measured CTDI vol were calculated for each scanning condition. For task-based image quality analysis, task transfer function, noise power spectrum, and system performance were evaluated. Streak artifacts, image noise, and contrast-to-noise ratio were quantified using a chest phantom.
Results:
With the Ag additional filter, mean energy was 22% higher and the CTDI vol was approximately 30% lower than those without the Ag additional filter. The task transfer function and noise power spectrum with the Ag additional filter were lower than those without the Ag additional filter. The system performance with the Ag additional filter was similar to that without the Ag additional filter. The Ag additional filter reduced streak artifact near the lung apex and image noise in the lung fields. The contrast-to-noise ratio was slightly higher with the Ag additional filter than that without the Ag additional filter.
Conclusions:
The output dose and spatial resolution with the Ag additional filter were lower than those without the Ag additional filter. However, this filter helped reduce the radiation dose, image noise, and streak artifacts, particularly when scanning at ultralow doses.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography