Related Experiment Video
Updated: Jul 19, 2025

Preparation, Administration, and Assessment of In Vivo Tissue-Specific Cellular Uptake of Fluorescent Dye-Labeled Liposomes
Published on: July 30, 2020
Box-Behnken Design-Aided Validation and Optimization of a Stability-Indicating Reverse Phase-HPLC Method for the
Gautam Kumar1,2, Prashansha Mullick3, Krishnadas Nandakumar1
1Department of Pharmacology, Manipal College of Pharmaceutical Sciences, Manipal Academy of Higher Education, Manipal, Karnataka 576104, India.
Abstract:
Stability indicating a reverse-phase HPLC analytical method for the quantification of tamoxifen citrate (TMX) in the bulk and lipidic nano-vesicles (LNVs) was developed. The optimized method was validated according to the ICH Q2 (R1) guidelines by following a three-factor interaction Box-Behnken design using Design-Expert® software. The responses measured at 236 nm were retention time (Rt), peak area, tailing factor (TF) and the number of theoretical plates. TMX was eluted best using the Luna® C18 LC Column along with a mobile phase of methanol (MeOH) and ammonium acetate buffer (AAB pH 4.5) 80:20 v/v mixture at 25 ± 2°C temperature. The currently developed method was linear in 100-5,000 ng/mL range with a detection limit of 4.55 ng/mL and a quantification limit of 13.78 ng/mL. The optimized method was utilized to evaluate the stability of TMX in different stress conditions by performing forced degradation studies. The results from the degradation study stipulated that on exposure to various stressors namely acid, alkali, oxidative, thermal and UV light, the TMX did not show considerable degradation except for UV light exposure. Further, the method was successfully used for the quantification of TMX in LNVs.