Related Experiment Video
Updated: Jul 19, 2025

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Denali Inferior Vena Cava Filter Retrieval: Complications and Success Rates
Purpose:
The present study was to evaluate the outcomes of Denali filter retrieval.
Materials And Methods:
We retrospectively reviewed 143 patients who received Denali filter insertion from September 2015 to April 2020. Ninety-seven patients who required removal of the filters were include in this study. Filters were retrieved with either standard or advanced techniques. Venography before and after retrieval was obtained to evaluate technical success, complications and duration of filter insertion.
Results:
All 97 filters were retrieved successfully without complications. Ninety-two (94.8%) were retrieved with standard technique and 5 filters (5.2%) required the advanced technique. There were two cases with a filter angle greater than 15 degrees. Inferior vena cava penetration was shown in 17 patients (17.5%) on venography but was not associated with contrast media extravasation after filter removal.
Conclusion:
The Denali filter showed a high rate of successful retrieval without complications. This study adds value to previous studies and trials showing that the Denali filter is a reliable and safe filter that can potentially improve retrieval rates, with increasing use of this device.
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications
Cardiac Catheterization I: Pre-Procedure Overview
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care