Related Experiment Video
Updated: Jul 19, 2025

Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Sleep apnea severity in patients undergoing atrial fibrillation ablation: Home sleep apnea-test and polysomnography
Nobuaki Tanaka1, Masato Okada1, Koji Tanaka1
1Cardiovascular Center Sakurabashi-Watanabe Hospital Osaka Japan.
Background:
Sleep apnea (SA) is highly prevalent and should be treated in patients referred for catheter ablation (CA) of atrial fibrillation (AF). Watch-type peripheral arterial tonometry (WP) for home SA testing has demonstrated a high correlation of the apnea-hypopnea index (AHI) with Polysomnography (PSG), but the evidence of its accuracy in AF patients is not adequate.
Methods:
This study was conducted under a retrospective, single-center, observational design. We included 464 consecutive AF patients (age 65 ± 11 years, 76.5% male, 45.0% paroxysmal-AF) who received both WP and PSG during the periprocedural period of the CA. We compared the AHI using the WP (WP-AHI) to that using PSG (PSG-AHI).
Results:
The WP-AHI was 25.9 ± 12.7 and PSG-AHI 31.4 ± 18.9 (r = .48). Among 325 patients with a WP-AHI < 30, 116 (35.7%) exhibited a PSG-AHI ≥ 30. Only 12.5% of the patients were indicated for continuous positive airway pressure (CPAP) treatment only by the WP-AHI, while 70.9% were indicated for CPAP by the PSG-AHI according to the Japanese health insurance system. The best cut-off value of the WP-AHI was 18.1 to predict a PSG-AHI ≥ 20 with an area under the curve of 0.72 (95% confidence interval, 0.67-0.76).
Conclusions:
The WP-AHI and PSG-AHI were weakly correlated in AF patients receiving CA. About one-third of the patients with moderate SA using the WP was diagnosed with severe SA evaluated by PSG. The majority required PSG for the CPAP indication.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation